Operation
37
Here’s an example of how to determine the attenuation
correction factor that needs to be set in the FieldMaxII-
TOP meter: If a 1 W laser beam is focused through an
attenuator that has an attenuation factor of 50%, then, to
get a true laser measurement value, the correction factor
in the FieldMaxII-TOP instrument needs to be set to 2. In
other words, since only half the power of the beam is
transmitted through the attenuator, the measured result
must be doubled to obtain a true laser measurement.
The Atten button is used to initiate several activities:
• Toggle the state of Attenuation Correction mode, if
the button is pressed for less than one second.
• Enter Edit mode and select the attenuation factor, if
Setup/Local is pressed beforehand.
• View the attenuation factor, if the button is pressed
for one second or more.
Trig
After pressing the Setup/Local button, the Trig button is
used to enter Edit mode and select the trigger level
parameter that will be edited. If an edit cycle has not been
initiated, pressing the Trig button will display the trigger
level parameter. Trigger has a range of 2 to 20% of full
scale. Refer to “Internal Triggering Mode” on page 48
for details of the trigger function